_RoHS环保测试仪作业指导手册

时间:2022-01-08 13:17:06 浏览量:

RoHS环保测试仪作业指导书

RoHS environmental testing working instruction

班级:姓名:学号

一.目的purpose:

为正确使用仪器,避免因不当使用导致仪器损坏,确保仪器使用寿命及测试准确度。

For the proper use of equipment, avoid the improper use of cause equipment damage, ensure the service life of the apparatus and testing accuracy.

二.范围Range:

用于进料、制程、出货阶段之( 原物料、辅材、半成品、成品等 )环境有害物质检测分析。

For feed, process, ship stage (raw materials, semi-finished products, finished products such as auxiliary, environmental hazardous substances testing analysis).

三.权责Responsibilities:

3.1. 仪校人员负责操作指导书制定。

3.1. Instrument school personnel responsible for operating instructions.

3.2. 测试人员负责仪器保养与维护。

3.2. The test personnel responsible for the equipment maintenance and maintenance.

3.3. 测试人员必须为接受过相关仪器操作培训合格之品保课人员。

3.3. The test must have received relevant equipment operation training of qualified quality assurance staff.

四.仪器开机与关机顺序Instrument boot and shutdown sequence:

4.1. 开机顺序: 开稳压电源开测试仪(绿灯亮) 开电脑主机打开测试程序软件。

4.1 boot sequence: regulated power supply testing instrument (green light) to open the computer host open test software.

4.2. 关机顺序: 退出测试程序软件关电脑主机关测试仪关稳压电源。

4.2. The shutdown sequence: exit test software to turn off the computer shut off power supply tester host.

五.指示灯说明Indicator description:

5.1. 绿灯-[电源指示灯] :打开仪器电源此灯亮。

5.1 - [green power indicator]: open the instrument power the lights.

5.2. 黄灯-[运行指示灯]:仪器测试运行时此灯亮,测试结束此灯自动熄灭。

5.2 yellow lights - [running indicator]: instrument test running the lights, the end of the test the lamp is automatically extinguished.

六.操作流程Operation process:

6.1. 每次开机时,首先需对仪器进行预热30分钟,具体预热步骤如下:

6.1. When starting up every time, the apparatus for preheating for 30 minutes, the preheating step as follows

6.1.1 apply (plastic sample) into the testing table click [sample] to view the sample placed on the toolbar

[参数设置]设置预热时间(1800) 秒点击[测塑料中Cr Cl]测量曲线点击开始键

RoHS环保测试仪作业指导手册

[] 输入样品

[parameters] preheat time (1800 seconds) click on the [Cr Cl] measurement measuring plastic curve click Start button input samples[]

信息(任意输入), 点击确定预热测量开始(黄灯亮),待1800秒测试时间终止,本次预热

即完成。

Information (arbitrary input), click OK to preheat the measurement start (yellow light), to be 1800 seconds test time expires, the preheating is finished.

6.2. 每次开机后或(仪器测试程序软件关闭后再打开)需测量产品前,必须先用纯银校正片作初始化校正,在此动作顺利通过情况下,方能继续下一步操作。具体初始化步骤如下:

6.2. After each reboot or (instrument testing software is closed and then opened) for measuring product, must first use silver correction piece as initialization correcting, this moves smoothly through the case, to continue the next operation.The initialization procedure is as follows:

“1105”道,否则需重新进行初始化达到要求为止。

6.2.1 select the toolbars [parameters] of (initialization), prompted by the school in sterling silver. Click [OK] to start the initialization initialization, after the success of the state bar of [peak channel] need to display as "1105", or the need to initialize meet the required date.

6.3. 样品测试操作步骤如下:

6.3 sample test steps are as follows:

6.3.1 measurement time settings: select the toolbars [parameters] of (measure time), appears in the dialog box input time (units / sec), click on the [OK] set complete.(setting time is generally 200 seconds, if the result is not accurate when may be appropriate to extend the time measurement)

6.3.1 measurement time settings: select the toolbars [parameters] of (measure time), appears in the dialog box input time (units / sec), click on the [OK] set complete.(setting time is generally 200 seconds, if the result is not accurate when may be appropriate to extend the time measurement)

RoHS环保测试仪作业指导手册

],按系统提示输入样品信息后,点击[OK]后测试开始,运行指示黄灯亮, 若需停止

RoHS环保测试仪作业指导手册

测试,点击工具栏停止键[]即可,测试完成黄灯自动熄灭。(样品细微时放入测试杯中再进行测试,测试运行中可点击[谱图]查看测试状况)

6.3.3 test:the measured object in test bench, click on the [sample] view objects, adjust the object position, until the show clearly the sample surface.Click on the toolbar to start

RoHS环保测试仪作业指导手册

[], prompted by the input sample information, click on the [OK] after the test, operation

RoHS环保测试仪作业指导手册

indicating light, if it is necessary to stop the test, click the toolbar button stop [], test complete yellow lamp is automatically extinguished.(fine when placed in the test sample cup for testing, test run can click [spectrum] view test status)

若无需保存只打印测试报告,测试完成后点击(分析报告打印)即可。

If no need to save the only print test reports, after the test is completed click (analysis report print)

6.4. 仪器校正:选择绿色塑胶校正片(ERM -EC681K)按6.3内容执行,测试完成查看测试报告并对照校正片标准值,容许偏差在±10PPM左右,若偏差过大,需联系仪器生产厂商进行处理。

6.4 calibration: choose green plastic correction sheet (ERM - EC681K) according to the 6.3 elements of the implementation, testing complete view test report and control the correction sheet standard value, allowable deviation is in the range of about 10PPM, if the deviation is too large, need to contact the equipment manufacturers processing (为使仪器测试准确,需不定期用校正片对测试参数进行校对,确保仪器之准确性)(apparatus for making accurate test, need not regularly using the correction sheet of test parameters on proofreading, ensure that the instrument accuracy)

七.注意事项Matters needing attention::

7.1测试工作台及仪器保持洁净,如有杂物灰尘可用干软布或酒精棉球擦拭干净。

7.1 test bench and apparatus to keep clean, if there is debris dust free dry soft cloth or alcohol cotton to wipe clean

7.2测试物件放置工作台时注意不可使尘粒掉入其中,以防污染工作台窗口导致测量不准确。

7.2 test objects placed a careful not to fall into the dust, to prevent the pollution of workbench window lead to inaccurate measurements

7.3测试物件放置时需轻拿轻放,以免工作台窗口薄膜破损,样品掉入仪器内损坏仪器。

7.3 test objects placed to light-light, lest the workbench window film damage, damage the instrument within the apparatus samples fall into.

7.4短时间不使用仪器,勿须关掉仪器电源,以免重新打开电源又须进行仪器预热。

7.4 short time without the use of instruments, do not turn off the power equipment, so as not to reopen the power supply must be instrument preheat

7.5开机后长时间(2-3小时)未使用仪器时,若需做测试须进行初始化。

7.5 boot after long time (2-3 hours) without the use of instruments, if it is necessary to do tests should be initialized.

7.6测试数据疑不准确时,可重新测试并适当延长测试时间,提高仪器分析准确性。

7.6 The test data that is not accurate, can test and extend the testing time, improve the

accuracy of instrument analysis.

7.7所有测试时间设定均以(秒)为单位。

7.7 all time in test set (s) for the unit

7.8样品细微未满足测试要求份量、软件界面[记数率]未达到1000-8000时,可适当延长测试时间。

7.8 In samples of fine not meet test requirements weight, software interface [count rate] not to 1000-8000, may be appropriate to extend the time of test.

7.9工作室应保持室内温度在20±5°为宜,相对湿度在70%以下为佳,保持工作室干净。

7.9 studio should keep the indoor temperature at 20 + - 5 DEG is appropriate, relative humidity below 70% is better, to keep the studio clean

7.10勿频繁开关仪器,否则电流容易冲击硬件设施,雷电天气应关闭仪器。

7.10 do not frequent switch apparatus, otherwise easy to impact current hardware facilities, lightning weather closing apparatus.